Viresolve Cartridge Filter

Description:
Viresolve NFR Cartridge Filter 20 in. Code 7
Trade Name:
Viresolve
Qty/Pk:
Applications:
Retrovirus removal
Key Applications:
Protein Purification
Hold-up Volume, mL:
Configuration:
Cartridge
Chemistry:
Polyethersulfone
Cartridge Code:
Code 7 (2-226) O-rings w/locking-tabs and spear
Filtration Area, m²:
0.854
Maximum Differential Pressure, bar (psid):
Forward: 5.5 (80) @ 25 °C. Reverse: 3.4 (50) @ 25 °C
Oxidizable Substances:
Meet the requirements of the USP Oxidizable Substances Test after a water flush of: 8 L
Cartridge Nominal Length, in (cm):
20 (50)
Process Volume, L:
1200–2400
Viral Clearance Device Type:
NFF Filters
Packaging:
Double Easy-Open bag
Sterilization:
After wetting, may be autoclaved for 3 cycles of up to 60 min at 125 °C,
using liquid cycle, slow exhaust.
Bacterial Retention:
≥6 LRV for Φ6 (78 nm) at a challenge of 107 pfu/cm2
Seal Material:
Silicone
Gravimetric Extractables:
Extractables level after a 10 min 1.5 Lpm/ft2 flush, after 24 hrs in ASTM® Type 1 reagent-grade water at controlled room temperature: ≤ 70 mg
Air Diffusion at 23 °C:
≤32 mL/min @ 3.45 bar (50 psig) in water
Product Name:
Viresolve Cartridge Filter
Bacterial Endotoxins:
Aqueous extraction contains ≤0.5 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) Test
Device Material:
Polypropylene