Millistak+® Lenticular Disc Filters
Description:
Millistak+® Lenticular Disc Filters, CE20 media series, 16 in. diameter, 16-cell, Neoprene seal
Trade Name:
Millistak+
Qty/Pk:
1
Applications:
For clarification of serum, plasma, vaccines, cell culture or other fluids.
Key Applications:
Laboratory Filtration
Conductivity, µS/cm:
8–10 umho/cm after a pure water flush of 5 liters per ft2 (50 L/m2) of media surface area per autoclaved (1 cycle of 30 minutes at 123 °C) 12" 16-stack device
Filter Grade:
20CE
Configuration:
Stack Disk
Chemistry:
Cellulose
Filtration Area, m²:
3.47
Maximum Differential Pressure, bar (psid):
Forward: 2.07 (30) @ 25 °C; 1.03 (15) @ 80 °C
Diameter, cm (in):
40.64 (16)
Height, cm (in):
27.6 (10.88)
Oxidizable Substances:
The effluent of an autoclaved (1 cycle of 30 minutes @ 123 °C) stacked device is negative after a WFI flush of 10 liters per ft2 (100 L/m2) of media surface area
Wettability:
Hydrophilic
TOC, ppb:
1370–2120 post autoclave (1 cycle of 30 minutes at 123 °C) and pure water flush of 5 liters per ft2 (50 L/m2) of media surface area
Sterilization:
May be steam sterilized for 3 steam-in-place (SIP) cycles of 30 minutes @ 123 °C
Sanitization:
May be hot water sanitized for 20 cycles using purified water @ 80 °C for 30 minutes
Seal Material:
Neoprene
Product Name:
Millistak+® Lenticular Disc Filters
Stack Design:
16 in. diameter, 16-cell
Metals:
- Sodium: 3.45 - 4.76 mg/ft2 of Media
- Calcium: 2.56 - 2.74 mg/ft2 of Media
- Potassium: 0.17 - 0.91 mg/ft2 of Media
- Magnesium: 1.19 - 2.03 mg/ft2 of Media
- Iron: <0.1 mg/ft2 of Media
- Lead: <0.1 mg/ft2 of Media
- Aluminum: <0.1 mg/ft2 of Media
- Titanium: <0.1 mg/ft2 of Media
- Chromium: <0.1 mg/ft2 of Media
- Manganese: <0.1 mg/ft2 of Media
- Cobalt: <0.1 mg/ft2 of Media
- Nickel: <0.1 mg/ft2 of Media
- Copper: <0.1 mg/ft2 of Media
- Zinc: <0.1 mg/ft2 of Media
Bacterial Endotoxins:
Aqueous extraction contains <0.25 EU/mL as determined by the Limulus Amebocyte Lysate (LAL) clot test technique
Device Material:
Polypropylene