Millipak-100 Filter Unit
Description:
Millipak-100 Filter Unit 0.1 µm Gamma Compatible 1-1/2 in. TC/TC
Trade Name:
Millipak
Key Applications:
Laboratory Filtration
Support Material:
Polycarbonate
Diameter, cm (in):
7.6 (3.0)
Sterility:
Non-Sterile
Packaging:
Double Easy-Open bag
Configuration:
Stack Disk
Vent Cap Material:
PVDF
Gravimetric Extractables:
2.5 mg
Filter Brand Name:
Durapore
Bubble Point at 23 °C:
≥4830 mbar (70 psig) air with water
Bacterial Retention:
Quantitative retention of 107 CFU/cm2 Brevundimonas diminuta (ATCC® 19146) per ASTM F838-83 methodology
Maximum Inlet Pressure, bar (psi):
5.2 (75) @ 25 °C
Process Volume, L:
50
Connections, Inlet/Outlet:
38 mm (1-1/2 in.) Sanitary Flange
Wettability:
Hydrophilic
Filter Pore Size, µm:
0.1
Maximum Differential Pressure, bar (psid):
Forward: 4.1 (60) @ 25 °C; 1.7 (25) @ 80 °C; 0.35 (5) @ 123 °C. Reverse: 0.7 (10) @ 25 °C
Capsule Type:
Liquid
Filtration Area, cm2:
500
Bacterial Endotoxins:
<0.5 EU/mL as determined by the LAL test
Filter Material:
Hydrophilic PVDF
Oxidizable Substances:
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≤200 mL
Length, cm (in):
11 (4.3)
Device Material:
Polycarbonate
Sterilization:
3 autoclave cycles of 90 min @ 123 °C; capable of 45 kilogray (4.5 Megarad) gamma exposure; not in-line steam sterilizable
Flow Rate, L/min:
3.8 @ 2.45 bar P