| Materials of Construction |
| Housing; Support Discs | Polycarbonate |
| Filter | Hydrophilic PVDF |
| Filtration Area, cm2 |
| 300 |
| Maximum Differential Pressure, bar (psid) |
| Forward: 4.1 (60) @ 25 °C; 1.7 (25) @ 80 °C; 0.35 (5) @ 123 °C. Reverse: 0.7 (10) @ 25 °C |
| Maximum Inlet Pressure, bar (psig) |
| 5.2 (75) @ 25 °C |
| Bacterial Retention (for Millipak 0.22 µm only) |
| Quantitative retention of 107 CFU/cm2 Brevundimonas diminuta (ATCC® 19146) per ASTM F838-83 methodology |
| Toxicity |
| Component materials meet the requirements of the USP Class VI Biological Test for Plastics. Millipak Filling Machine Filter units also meet the requirements of the USP General (Mouse) Safety Test and USP Elution (cytotoxicity) test. |
| Sterilization |
| 3 autoclave cycles of 90 min @ 123 °C; capable of 45 kilogray (4.5 Megarad) gamma exposure; not in-line steam sterilizable |
| Integrity Test, Bubble Point |
| 0.22 µm units | ≥3450 mbar (50 psig) air with water |
| 5 µm units | ≥138 mbar (2 psig) air with water |