| Per 10-inch Cartridge |
| Materials of Construction |
| Filter | Hydrophobic PTFE |
| Supports | Polypropylene |
| Cage and End Caps | Polypropylene |
| Core | Polysulfone |
| O-ring | Silicone |
| Filtration Area, m2 | 0.66 |
| Maximum Differential Pressure, bar (psid) | Forward: 4.1 (60) @ 25 °C; 0.35 (5) @ 145 °C |
| Virus Aerosol Retention | Retention of a worst case challenge of the model bacteriophage Φ X-174 was complete at challenge levels up to 107 to 1010 virus particles at 1400 sL/min (50 scfm) after 200 steam sterilization cycles of 30 minutes at 145 °C |
| Bacterial Growthrough | Prevention of bacterial growthrough by Brevundimonas diminuta for at least 21 days @ a challenge level ≥1 x 107 CFU/cm2 |
| Toxicity | Component materials meet the requirements of the USP Class VI Biological Test for Plastics. The cartridges also meet the requirements of the USP General (Mouse) Safety Test. |
| Sterilization | 200 forward SIP cycles of 30 min @ 145 °C |
| Integrity Test |
| Bubble Point | ≥965 mbar (14 psig) nitrogen with 70//30 % IPA/water |
| Nitrogen Diffusion | ≤60 mL/min @ 410 mbar (6.0 psig) in 70//30 % IPA/water |
| HydroCorr | ≤0.40 mL/min per 10-inch cartridge with water @ 690 mbar (10 psig) |