| Opticap XL10 | Opticap XLT10 | Opticap XLT20 | Opticap XLT30 |
| Filtration Area, m2 | 0.55 | 0.55 | 1.2 | 1.75 |
| Materials of Construction |
| Filter | Hydrophilic PVDF | Hydrophilic PVDF | Hydrophilic PVDF | Hydrophilic PVDF |
| Supports | Polypropylene | Polypropylene | Polypropylene | Polypropylene |
| Structural Components | Polypropylene | Polypropylene | Polypropylene | Polypropylene |
| O-ring | Silicone | Silicone | Silicone | Silicone |
| Vent/Drain | 1/4 in. hose barb with double silicone O-ring seal |
| Maximum Differential Pressure, bar (psid) | Forward: 5.5 (80) @ 25 °C; 1.1 (25) @ 80 °C. Reverse: 3.5 (50) @ 25 °C, intermittent |
| Bubble Point | ≥3450 mbar (50 psig) air with water |
| Air Diffusion | ≤10.8 mL/min @ 2.75 bar (40 psig) in water | ≤10.8 mL/min @ 2.75 bar (40 psig) in water | ≤21.6 mL/min @ 2.75 bar (40 psig) in water | ≤32.4 mL/min @ 2.75 bar (40 psig) in water |
| TOC/Conductivity | Autoclaved filter meets the WFI requirements of USP <643> for Total Organic Carbon and USP <645> for Water Conductivity after a WFI water flush of: 16 liters at 25 °C | Autoclaved filter meets the WFI requirements of USP <643> for Total Organic Carbon and USP <645> for Water Conductivity after a WFI water flush of: 16 liters at 25 °C | Autoclaved filter meets the WFI requirements of USP <643> for Total Organic Carbon and USP <645> for Water Conductivity after a WFI water flush of: 32 liters at 25 °C | Autoclaved filter meets the WFI requirements of USP <643> for Total Organic Carbon and USP <645> for Water Conductivity after a WFI water flush of: 48 liters at 25 °C |
| Bacterial Retention | Quantitative retention of 107 CFU/cm2 Brevundimonas diminuta ATCC® 19146 per ASTM methodology. |
| Bacterial Endotoxins | <0.25 EU/mL as determined by the LAL test per 10 inch cartridge |
| Oxidizable Substances | Will meet the USP Oxidizable Substances Test requirements after a water flush of ≤1.5 L | Will meet the USP Oxidizable Substances Test requirements after a water flush of ≤1.5 L | Will meet the USP Oxidizable Substances Test requirements after a water flush of ≤3 L | Will meet the USP Oxidizable Substances Test requirements after a water flush of ≤4.5 L |
| Sterilization | 3 autoclave cycles of 60 min @ 126 °C; not in-line steam sterilizable |
| Toxicity | Component materials meet the criteria of the USP <88> Reactivity Test for Class VI Plastics. These products meet the requirements of the USP <88> Safety Test, utilizing a 0.9% sodium chloride extraction. |
| Good Manufacturing Practices | These products are manufactured in a facility which adheres to FDA Device Good Manufacturing Practices. |
| Non-Fiber Releasing | Component materials meet the "non-fiber releasing" criteria as defined in 21 CFR 210.3 (b) (6). |
| Indirect Food Additive | All component materials meet the FDA Indirect Food Additive requirements cited in 21 CFR 177–182. |