> Último artículo añadido:
0 artículo/s

Amicon Ultra-2 mL Pre-Launch Centrifugal Filters for DNA Purification and Concentration

With their vertical membrane design, Amicon Ultra 2 mL filters deliver great performance and lower spin times
 

  • Concentrates 2 mL down to 15-70 µL
  • 98% sample recovery
  • Fast processing time - 10 to 30 minute spin times
  • Engineered dead stop
  • Reverse spin capability provides consistent recoveries

Pre-launch Device Disclaimer

This pre-launch product is in the final stages of both product and process development, and validation activities are ongoing. The information presented is based on preliminary development results. Device performance in specific applications may be different.

Applications

  • Concentration of biological samples containing antigens, antibodies, enzymes, nucleic acids, or microorganisms
  • Purification of macromolecular components found in tissue culture extracts or cell lysates, primer removal, linkers, or macromelecular labels from a reaction mix, and protein removal prior to HPLC
  • Desalting, buffer exchange, and protein dialysis

Not sure if you found the right Amicon for your application?
» Search with our new Amicon Ultra Selector tool.

For Protein Purification and Concentration click here

Performance – DNA Concentration

The Amicon Ultra-2 30K device provides the best balance between PCR recovery and PCR primer removal for double-stranded DNA for base pairs ranging from 137 to 1159.

Swinging Bucket Rotor 35° Fixed Angle Rotor
Amicon Ultra
Device
PCR Product
(base pairs)
PCR Primer (bases) PCR Recovery
(%)
PCR Primer
Removal (%)
Final Volume
(μL)
PCR Recovery (%) PCR Primer
Removal (%)
Final Volume
(μL)
30K 137 10 92 100 37 93 100 27
20 94 93 39 94 95 31
48 95 71 36 93 78 31
100K 137 10 60 100 33 56 100 16
20 50 100 33 41 100 17
48 50 92 35 39 91 16
Spin conditions: Swinging bucket rotor, 4,000 x g for 40 min; fixed angle rotor, 7,500 x g for 15 min; room temperature, 100 μL PCR diluted to 2,000 μL starting volume, n=3.

Desalting or Diafiltration

Desalting, buffer exchange or diafiltration are important methods for removing salts or solvents in solutions containing biomolecules. The removal of salts or the exchange of buffers can be accomplished in the Amicon Ultra-2 mL device by concentrating the sample, then reconstituting the concentrate to the original sample volume with any desired solvent. The process of "washing out" can be repeated until the concentration of the contaminating microsolute has been sufficiently reduced. See example below.

Amicon Ultra-2 mL Centrifugal Filters for Protein Purification and Concent


Contactos

Queremos tener noticias suyas. Comunique con uno de los siguientes departamentos o díganos su opinión haciendo clic en "Sugerencias y Comentarios".