Fast-Trap Adenovirus/Lentivirus Purification and Concentration Kit
Frequently Asked Questions
Q. I work with a different pseudo/serotype of Lenti/Adenovirus (other than VSV-G for Lentivirus, or serotype 5 for Adenovirus), will they work with this kit?
- These kits have been optimized and tested using a VSV-G pseudotype Lentivirus, Adenovirus serotype 5 and AAV serotype 2. Other similar viruses may work, but have not been evaluated for performance. We have heard from customers that the AAV kit will work with AAV serotypes 1 and 7 as well.
Q. How many T-flasks provide the appropriate amount of virus for use of these kits?
- For Adenovirus, the equivalent of between 2-20 T-150cm2 flasks generally yields sufficient amounts of virus for use with this kit. However, the titer of the total number of viral particles should be used as the basis for the amount of virus used with the Fast-Trap kit (e.g. depending on cells, specific virus, processing conditions, etc. the amount of virus generated per T-flask will vary considerably. Therefore, some customers may reach capacity with 10 flasks while others may reach capacity with 20.
For Lentivirus, the equivalent of between 2-8 T-75 flasks generally yields sufficient amounts of virus for use with this kit. Again, the titer of the total number of viral particles should be used as the basis for the amount of virus used with the Fast-Trap kit.
For AAV, the kit can handle virus produced from up to 1025 cm2 cell culture surface area. Typically this limit is equivalent to 13 T-75 flasks but again, the titer of the total number of viral particles should be used as the basis for the amount of virus used with the Fast-Trap kit.
Q. My Adenovirus and Lentivirus sample volume is larger than 50ml. Is it possible to process this sample?
- Yes. It is possible to load sample in sequential steps as long as capacity of the device is not exceeded. Also, make sure to change out the 50ml collection tube with an empty tube to ensure the filtrate volume does not exceed 50ml.
Q. Is Benzonase treatment of my sample required for these kits?
- No. However, optimal results are achieved when using a nuclease pre-treatment step prior to purification. Faster flow rates have been observed with samples treated with Benzonase.
Q. Why does my flow rate slow down after adding my virus sample?
- As virus binds to the membrane there is less space available in the pores of the membrane thereby decreasing the rate at which liquid can flow through the membrane. A significant decrease in flow rate may be observed if too much virus is loaded onto the device. Purification of virus titers above recommended capacity is best performed by splitting the challenge between multiple purification devices.
Q. I don’t experience a gradual decrease in flow rate; rather the flow rate is constantly slow. Why could this be happening?
- Ensure the vacuum pressure is at least 15” Hg for the device to function properly. Also, make sure the Fast-Trap purification device and 50ml conical tubes are tight and securely fastened.
Q. My clarification filter clogged (Steriflip-HV) why did this happen, what can I do?
- A crude virus stock with too much cellular debris will clog the clarification unit. This can be prevented by performing the post-harvest centrifugation step (after freeze/thaw for Adenovirus) until no debris is visible in the supernatant. If a clarification unit clogs, the crude virus sample can undergo an additional centrifugation step followed by clarification with another Steriflip-HV unit.
Q. Do I really need to dilute my virus sample in binding buffer?
- Yes. Diluting the starting virus sample in the 10X binding buffer allows for optimal binding of the maximum amount of virus particles to the membrane.
Q. Will a second elution result in additional recovery?
- A second 3ml elution typically results in an extra 1-5% recovery.
Q. Is my virus sample sterile after purification/concentration?
- No. Sterilization of final concentrated sample may be performed using a Millex-GP Filter Unit.
Q. Do I need to perform buffer exchange after purification?
- The purified virus eluted from the Fast-Trap device is in a high salt buffer. Exchanging purified virus into the optimal buffer for storage and/or downstream processing is recommended. An example of one storage buffer is: 20mM Tris-HCl pH 8.0, 250mM NaCl, 5% sorbitol, and 0.001% PF68 (pluronic acid)
Q. My virus recovery is below 40%. What went wrong?
- Titer of starting samples exceeded maximum capacity of Fast-Trap device (Adenovirus capacity <1x1013 total particles/1x1011 infectious particles, AAV capacity is 1x1012 total viral particles, and Lentivirus capacity < 2x108 infectious particles)
Reducing the sample volume below 500µl for Adenovirus or 200µl for Lentivirus in the Amicon Ultra during buffer exchange/concentration may cause the virus to aggregate resulting in decreased recovery of infectious virus particles.
Also, ensure that during the elution procedure 5 drops of elution buffer have filtered through the membrane prior to the 5 minute incubation. This allows maximum interaction between the elution buffer and the virus adsorbed onto the membrane. Failure to follow this procedure may result in decreased virus recovery.
Purification of crude virus containing sucrose/glycerol (sometimes used by customers to stabilize Lentiviral samples) can result in low virus recovery.
Q. All this connecting and disconnecting of tubes is inconvenient. Is there another option?
- Yes. Millipore provides an optional funnel attachment (Catalogue Number: SC50FL025) for the top of the Fast-Trap purification device that allows equilibration, sample, wash, and elution buffers to be added directly to the purification device.
Q. What specific virus has the Fast-Trap kits been tested on?
- The Adenovirus kit was developed using Adenovirus Serotype 5, which is the most commonly used Adenovirus for virus purification. We do not know whether or not the kit will work the same with other serotypes of Adenovirus. It may, but it has not been tested on other serotypes.
The Lentivirus kit was developed using the VSV-G Serotype of Lentivirus, which is the most commonly used Lentivirus type for virus purification. We do not know whether or not the kit will work the same with other pseudotypes of Lentivirus.
Tha AAV kit was developed using AAV serotype 2. We have heard from customers that it works with AAV serotypes 1 and 7 but we have not tested this ourselves.
Q. Depending on how much virus I am trying to purify, the flow rate slows down.
- Flow rate is expected to decrease when using high virus titer. Do not exceed device capacity of approximately 1x1013 total virus particles for the Adenovirus kit, 1x1012 total viral particles for AAV and 2x108 infectious particles for the Lentivirus kit.
Q. Can the virus be stored in the elution buffer - what are the storage conditions? Does anything have to be added?
- The elution buffer contains high (>1M) salt/Tris-HCl (pH 8), and should be buffer exchanged (w/2 spin protocol) with the Amicon Ultra before using for their application or addition of glyerol for long-term storage. If it is not buffer exchanged, the tube could form salt crystals and might not freeze properly. Also, if kept in the elution buffer, their downstream applications (i.e. with cells) might be adversely impacted.
| 1. Examples of "exchange" buffer could be: | ||
| ||
2. Customer can also try the buffer that they typically use for an "exchange" buffer. | ||
Q. Do you know how the Fast Trap Lentivirus kit compares to PEG precipitation?
- The Fast-Trap Lentivirus was compared side-by-side with the System Biosciences PEG-it precipitation: (Catalogue Number LV810A-1):
- Both products had similar recovery (~57% of input 2.2x107 infectous viral particles).
- However, the purity (using gel electrophoresis and Western Blotting) was very poor with the PEG-it when compared to the Fast-Trap
- Although there is a lot of hands-off time with the PEG-it, it took over 12.5 hours to complete, vs. < 2 hours with the Fast-Trap.
Q: Which membrane is used in the fast trap devices?
- New membranes were developed for these kits. For Lentivirus and Adenovirus the membrane is an anion exchanger. For Adeno-Associated Virus the membrane is a cation exchanger.
Q: Why is there a difference in Amicon Ultra molecular weight cutoff depending on the virus type?
- The AU cut offs by virus are the ones recommended in the virus purification tech brief for the specific virus type.

