Specifications

The LC-Pak cartridge was extensively validated to ensure that the water quality at the outlet matches the most demanding requirements.

Parameter Specification of LC-Pak ultrapure waterComments
HPLC Gradient Test - Absorbance of highest eluted peak At 210 nm < 0.006 AU
At 254 nm < 0.002 AU
Concentration of 60 mL water at 1 mL/min prior to elution
HPLC Gradient Test - Absorbance of highest eluted peakAt 210 nm < 0.003 AU
At 254 nm < 0.001 AU
No water preconcentration
Optical properties: absorbance in UV rangeUV 200 nm < 0.05 AU
UV 205 nm < 0.01 AU
UV 210 nm < 0.01 AU
UV 254 nm < 0.005 AU
Fluorescence as quinine At 254 nm < 1 ppb
At 365 nm < 1 ppb

Compliance with suitability for LC/MS: Reserpine test
No peak higher than 10 ppb
Reserpine at 609 m/z in ESI +
Residue after evaporation < 0.0001 % w/w Test performed as specified in ISO 3696 procedure

An LC-Pak validation guide is available should you wish to review the details of the validation report.


Top »