Validating the Guava ViaCount Cell Dispersal Reagent Assay

Lit No:MK46000560
Year:2003


Accurate cell counts and viability assessments of homogeneous cell suspensions can be accomplished by a variety of methods. The Guava ViaCount assay provides cell count and viability data for cell suspensions from a variety of cell lines, quickly and easily. However, cell samples containing clumpy aggregates present a challenge by any method, manual or automated. We have developed the Guava ViaCount CDR assay to address this problem. The assay uses Guava ViaCount CDR Cell Dispersal Reagent to gently disaggregate clumps, and Guava ViaCount Reagent to stain cells for count and viability assessment. The assay was optimized and validated using CHO-K1 cells that were adapted to suspension culture in serum-free medium. Significant improvement in the count accuracy was attained for this assay, with good linear correlation to hemacytometer counts (R2 > 0.93) over a broad range of cell densities (5 x 105 – 1 x 107 cells/mL) and culture conditions (50 - 95% viable, 17 - 87% aggregation). Assay results were highly reproducible, with % CV for replicate cell count values ranging from 3 - 7% and % CV for replicate viability measurements of 1 - 2%.