Demonstrating the superiority of the Guava ViaCount Cell Counting/Viability Assay over the Trypan blue method

Lit No:MK80000410
Year:2009


In order to assess counting performance of the Guava ViaCount assay with mesenchymal stem cells, twelve cell samples were obtained and counted using both the ViaCount assay performed on the Guava PCA™ system and a trypan blue-based assay on a hemacytometer1. In addition, four of the samples were assayed after dilution to 50% of their initial concentration. Data for total cell counts are shown in Figure 1 along with linear regression statistics. In this study performed by Osiris Therapeutics, Inc., a strong correlation between the two assays was observed, as indicated by a slope of 1.02 and an R2 value of 0.985.