19. Error Messages
User Guide
XITXSP121P-A


TEST STATUS ERRORS
During the operation of the Integritest Exacta unit, certain error conditions are tested for and if found the test is terminated and an error status message is reported on the test progress screen. In addition, there is a more detailed text description of the error condition that appears in a message window (Figure 82). For further details refer to the Operations Log.

Figure 82: Example of an Error Message Box

Table 27: Status Error Messages

Status Error No.Status Window MessageShort Description Detailed Description
1MATH ERRORGeneral FailureGeneral error in operation Used to convey errors occurring during operation that include, improper set up parameters, computational errors, etc.
2INPUT ERROR Failed Precondition CheckError that occurs at the initiation of a test. The input parameters and the test & math dll‘s are checked for validity prior to initiating the test. This error condition will arise if a condition fails.
3OUTPUT ERRORFailed Postcondition CheckError that occurs at the completion of a test. The output termination parameters and are checked for validity at the completion of the test. This error condition will arise if a condition fails.
4INPUT TOO SMALLFailed Param Low Range CheckFailed minimum boundary value for the input value listed.
5 INPUT TOO HIGHFailed Param High Range CheckFailed maximum boundary value for the input value listed.
6NETWORK ERRORRPC FailureIn network communications the RPC (Remote Protocol Control) layer is the set of functions that allow an IT Exacta hardware control software to communicate with a remote User Interface (either in an IT Exacta or on another computer). This error message is generated when there is a failure in this communications link and the two are not connected.
7NETWORK UNAVAILABLERPC Service Not AvailableThe Remote Communications Control package is being used by another computer and unavailable at this time.
8 GENERAL ERROR INSUFFICIENT GAS PRESSURETest Execution General FailureGeneral fault condition associated with an error in the test protocol. Generally caused by incorrect starting values being transmitted to the hardware for either test operation or system configuration. See the operations log for details. Insufficient inlet gas pressure to run a test.
9UNDEFINED TESTTest Procedure Not FoundAt the time of execution, the test being asked for is checked against the tests stored within the internal database to ensure that the test is properly defined and available for use. This error indicates that the requested test procedure is not found in the database test catalogue.
10UNDEFINED TEST REVISIONTest Version Not FoundSame as above, only the test version number is the entity being checked.
11TEST IN USETest Already RunningError condition flagged when the requested test is already running on the IT Exacta hardware.
12 NO TESTNo Test AvailableError condition that occurs at the initiation of a test. If the test that is requested cannot be found this error is generated. This condition occurs if an archived test is used or if a stored test requests a revision number that is different from the test being requested.
13NO TESTRequested Test UnavailableError condition that occurs at the initiation of a test. If the test that is requested cannot be found this error is generated. This condition occurs if an archived test is used or if a stored test requests a revision number that is different from the test being requested.
14NO DATARequested Test Data UnavailableError condition that occurs from the data management function. The data being requested cannot be found in the database.
15GROSS LEAKLeak Detected During the test, several checks are made (at every data point) to determine the abrupt start of a gross leak. The system is aborted if a gross leak is found. A Gross leak can occur if tubing is ruptured etc. or also in a bubble point or enhanced bubble point test if the pressure is increased too high and the filter is completely blown dry.
16TEST TIME EXCEEDEDMax Time ExceededThe maximum test time has been exceeded.
17GROSS LEAK Max Loss ExceededDuring each pressure decay the total pressure drop is monitored, and if it exceeds half the test pressure this error is shown.
18CALIBRATEOut Of CalibrationThis error indicates that the transducer or pressure regulator is out of calibration when compared to the manufacturer specifications. A mechanical adjustment or replacement is required.
19USER TERMINATEDUser Aborted TestUser termination.
20 NO FILTERFilter Not FoundThe filter being requested is not found in the filter database. Both the filter catalogue number and revision number needs to match that being requested.
21NO HC INPUTSHC Inputs Not FoundSpecific to a HydroCorr test, the HydroCorr input values could not be found.
22SETUP INSTRUMENTSetup Inputs Not FoundSetup inputs such as test pressure, minimum bubblepoint, test time could not be found in the database.
23NO ERROR LOGClient Log ErrorThere has been an error in accessing the log file associated with the user interface.
24UNDEFINED USERUser Not FoundLog in error – the user name is not valid for this network

General Messages
Messages may appear during any instrument operation. Table 28 lists these messages with their cause and location.

Table 28: General Error Messages
Scope/ScreenMessage Cause
GlobalCould not read access values from Database. Using
default values.
Database Communication Error
GlobalCould not store access level changes on the server.Unable to Store the Access Levels in the Database
LoginUnable to connect to master serverN/A
Login Exacta service is unavailable on:Network Login failure
Login The Exacta service on this instrument is not currently availableThe server is not running or there is a network error
Global There was a problem printing to your printer.Printer Failure
GlobalAn instrument error has occurred - The current operation has been terminated - Please consult the event logs for more details.Generic Test Error
External Calibration
Dialog Box
A valid referee value is requiredAn Invalid Referee Value Was Entered
External Calibration
Dialog Box
The referee temperature specified is out of range.Invalid Temperature Value Entered
External Calibration
Dialog Box
The referee pressure specified is out of range.Invalid Pressure Value Entered
Test Progress Screen The test version you are trying to run is not available on ServerThe Test you tried to run is not valid for some reason
Test Progress Screen The test version you are trying to re-run is not available on ServerThe Test you tried to re-run is not valid for some reason
Test Progress Screen The raw data for the old test is missing from the database.The data for the Test you attempted to rerun could not be found on the server
Test Progress ScreenUnable to abort test Unable to abort the test for some reason
Test Progress Screen The instrument is not currently available.Instrument Not Available
Lock Out Form Are you absolutely sure that you want to unlock this instrument?You are about to unlock the instrument
Lock Out Form The instrument is not currently availableInstrument Not Available
Node Admin Changes have been made. Save them?Changes have not been saved
Utilities The instrument is not currently available.Instrument Not Available
HW Diagnostics Pressure setting invalid. Defaulting to 0Invalid Pressure Value Entered
Test Data Please enter a range of days where the end date is after the start date.Please enter a range of days where the end date is after the start date.
Test Data The instrument is not currently available.Instrument Not Available
Equipment Setup Changes have been made. Save them?Changes have not been saved
Equipment Setup Temperature override is invalid.Invalid Temperature Value Entered
Equipment Setup Local barometric pressure is invalid.Local barometric pressure is invalid.
Equipment Setup Temperature override is invalidInvalid Temperature Value Entered
Equipment Setup The instrument is not currently available.Instrument Not Available
Equipment Setup Error: Last Test Status = External Pressure Calibration Failure Code
Equipment Setup The instrument is not currently available.Instrument Not Available
Equipment Setup Error: Last Test Status =Pressure Calibration Failure Code
Equipment SetupThe instrument is not currently available. Instrument Not Available
Equipment SetupError: Last Test Status = Pressure Calibration Failure Code
Equipment SetupThe instrument is not currently available.Instrument Not Available
Equipment Setup Error: Last Test Status = Pressure Calibration Failure Code
ClientSetup Automatic Logout Time is InvalidAutomatic Logout Time is Invalid
ClientSetup Database error : Database Error during Load
ClientSetup Printer Error Return Code Printer Error
User Admin Changes have been made. Save them?Changes have not been saved
User AdminUnable to connect to master serverThe Server Software is not Running or the Network is Configured Incorrectly
Operator Input Return Database Error CodeDatabase Error during Load
Operator InputThe instrument is not currently available.Instrument Not Available
Operator Input You must enter a value for the following input:Error in Operator Input
Global No current form.Failed Attempt to Change Screen
Global Form stack overflow.Too Many Forms Loaded
Advanced Query Please enter a range of days where the end date is after the start date.Please enter a range of days where the end date is after the start date.
Access Level Changes have been made. Save them?Changes have not been saved
Filter Editor Changes have been made. Save them?Changes have not been saved
Filter Editor The following fields are invalid:Check the Fields Mentioned
Filter Editoris an invalid character Check the Fields Mentioned
Filter Editor Database error : Database Error during Load
Test Editor The filter revision update has been undone. The test cannot be edited until the filter version is brought up to date.Filter out of date
Test Editor The filter associated with this test is not the latest version.
The test cannot be edited unless the most current filter is in use. Update the filter attributes of this test with those from the current filter definition?
Filter out of date
Test Editor is an invalid character Check the Fields Mentioned
Test Editor Database error : Database Error during Load
Test Editor Changes have been made. Save them?Changes have not been saved
Test Editor The following fields are invalid:Check the Fields Mentioned