Guava Check Kit

Lit No.:MK46000080
Year:2007


The Guava Check Kit is used to verify the performance of your Guava PCA, Guava PCA-96, Guava PCA-96 AFP, Guava EasyCyte™, Guava EasyCyte Plus, or Guava EasyCyte Mini System by assessing counting accuracy and fluorescence detection using a standardized fluorescent bead reagent. The kit contains Guava Check Beads and Diluent. Running the Guava Check Procedure daily, before performing any assays, is important to assure that your Guava Systems provide reliable, accurate results. Guava Check Beads are provided as a concentrated fluorescent bead suspension of known concentration. To run the Guava Check Procedure, mix the bead reagent with Guava Check Diluent at the recommended dilution to yield a bead suspension at a specific target concentration. The Guava Check Software Module analyzes the results from three sequential data acquisitions of this bead suspension. The count results are displayed on the screen for comparison with the known target value. Particle counts that are outside the expected range are displayed in red. Count results, forward scatter and side scatter (optional on some systems) signal intensity, and mean fluorescence intensities are displayed. The averages and %CV values for these results are calculated and also displayed. These results and the instrument settings are saved to a spreadsheet file. The results from all Guava Check Procedures are appended to a file, providing a history of data that allows you to monitor the performance of your Guava System.

Click the PDF above for the full document.